Article Text
Abstract
Aims Recently, a new automated digital cell imaging analyser (Sysmex CellaVision DC-1), intended for use in low-volume and small satellite laboratories, has become available. The purpose of this study was to compare the performance of the DC-1 with the Sysmex DI-60 system and the gold standard, manual microscopy.
Methods White blood cell (WBC) differential counts in 100 normal and 100 abnormal peripheral blood smears were compared between the DC-1, the DI-60 and manual microscopy to establish accuracy, within-run imprecision, clinical sensitivity and specificity. Moreover, the agreement between precharacterisation and postcharacterisation of red blood cell (RBC) morphological abnormalities was determined for the DC-1.
Results WBC preclassification and postclassification results of the DC-1 showed good correlation compared with DI-60 results and manual microscopy. In addition, the within-run SD of the DC-1 was below 1 for all five major WBC classes, indicating good reproducibility. Clinical sensitivity and specificity were, respectively, 96.7%/95.9% compared with the DI-60% and 96.6%/95.3% compared with manual microscopy. The overall agreement on RBC morphology between the precharacterisation and postcharacterisation results ranged from 49% (poikilocytosis) to 100% (hypochromasia, microcytosis and macrocytosis).
Conclusions The DC-1 has proven to be an accurate digital cell imaging system for differential counting and morphological classification of WBCs and RBCs in peripheral blood smears. It is a compact and easily operated instrument that can offer low-volume and small satellite laboratories the possibilities of readily available blood cell analysis that can be stored and retrieved for consultation with remote locations.
- haematology
- morphological and microscopic findings
- cell differentiation
- leucocytes
- erythrocytes
Data availability statement
All data relevant to the study are included in the article or uploaded as supplementary information.