PT - JOURNAL ARTICLE AU - Moses, G C AU - Thibert, R J AU - Draisey, T F TI - Continuous-flow analysis of ammonia in perchloric acid supernate of blood or plasma using an ammonia-selective electrode. AID - 10.1136/jcp.31.12.1207 DP - 1978 Dec 01 TA - Journal of Clinical Pathology PG - 1207--1211 VI - 31 IP - 12 4099 - http://jcp.bmj.com/content/31/12/1207.short 4100 - http://jcp.bmj.com/content/31/12/1207.full SO - J Clin Pathol1978 Dec 01; 31 AB - A simple automated method for the estimation of ammonia in perchloric acid supernate of blood or plasma using an ion-selective electrode (Orion Ammonia-selective electrode, Model 95-10) is described. The reliability of the proposed method has been checked against an ion-exchange resin procedure, which has been chosen as a standard procedure. Regression equation and correlation coefficient for the proposed method are y = 0.7 X + 10 and 0.945, respectively, as compared with the chosen standard method. Within-run and between-run precision are 2.1% and 3.5% respectively. The average percent recovery is 97.5% and a tentative range is 13-73 microgram/dl (9-52 micronmol/l) ammonia nitrogen.