PT - JOURNAL ARTICLE AU - Van Der Gugten, Jessica Grace AU - Dubland, Joshua AU - Liu, Hua-Fen AU - Wang, Alex AU - Joseph, Christine AU - Holmes, Daniel T TI - Determination of serum aldosterone by liquid chromatography and tandem mass spectrometry: a liquid–liquid extraction method for the ABSCIEX API-5000 mass spectrometry system AID - 10.1136/jclinpath-2011-200564 DP - 2012 May 01 TA - Journal of Clinical Pathology PG - 457--462 VI - 65 IP - 5 4099 - http://jcp.bmj.com/content/65/5/457.short 4100 - http://jcp.bmj.com/content/65/5/457.full SO - J Clin Pathol2012 May 01; 65 AB - Aims Accurate serum aldosterone determination is critical to the screening and diagnosis of primary aldosteronism, the localisation of aldosterone producing tumours, and the investigation of other disorders of the renin-angiotensin system. Mass spectrometry offers a means to overcome problems with method-dependent bias between competitive immunoassays for aldosterone. The authors have developed a simple, sensitive and precise liquid–liquid extraction aldosterone method for the ABSCIEX API-5000 liquid chromatography and tandem mass spectrometry (LC-MS/MS) system.Methods Using d7-aldosterone internal standard, 500 μl of sample is extracted with 2500 μl of methyl tertbutyl ether followed by dry-down, reconstitution and LC-MS/MS analysis in ESI negative mode. Method validation was undertaken using standard approaches and comparison made against a commercial radioimmunoassay. Accuracy was assessed using EQA material with assigned aldosterone concentrations.Results The assay was linear up to 3420 pmol/l (LOQ=50 pmol/l, LOD<22 pmol/l). Total CVs were ≤5% for concentrations ≥120 pmol/l and 10% at the LOQ. Mean accuracy was 98.5% against GCMS assigned material.Conclusion The authors present a precise, sensitive and simple aldosterone method suitable for routine clinical use that requires no solid phase extraction or specialised ion sources.