Enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii based on enhanced chemiluminescence.
  1. C F Crouch
  1. Kodak Clinical Diagnostics Ltd, Pollards Wood Laboratories, Chalfont, St Giles, Bucks.

    Abstract

    AIMS--To evaluate the clinical performance of enzyme immunoassays for IgG and IgM antibodies to Toxoplasma gondii based on enhanced chemiluminescence. METHODS--Classification of routine clinical samples from the originating laboratories was compared with that obtained using the chemiluminescence based assays. Resolution of discordant results was achieved by testing in alternative enzyme immunoassays (IgM) or by an independent laboratory using the dye test (IgG). RESULTS--Compared with resolved data, the IgM assay was found to be highly specific (100%) with a cut off selected to give optimal performance with respect to both the early detection of specific IgM and the detection of persistent levels of specific IgM (sensitivity 98%). Compared with resolved data, the IgG assay was shown to have a sensitivity and a specificity of 99.4%. CONCLUSIONS--The Amerlite Toxo IgM assay possesses high levels of sensitivity and specificity. Assay interference due to rheumatoid factor like substances is not a problem. The Amerlite Toxo IgG assay possesses good sensitivity and specificity, but is less sensitive for the detection of seroconversion than methods detecting both IgG and IgM.