Multicentre evaluation of the Thrombotest International Sensitivity Index used with a steel ball coagulometer.
  1. A M van den Besselaar,
  2. R H Peters
  1. Department of Haematology, Leiden University Hospital, Netherlands.

    Abstract

    AIM: To compare the International Sensitivity Index (ISI) of the Thrombotest reagent used with a steel ball coagulometer (KC) to the ISI of the same reagent used with the manual (tilt tube) technique. METHODS: The study was carried out by eight laboratories using their own KC instrument and method of testing. All laboratories used the same batch of Thrombotest to determine the clotting times of fresh blood samples from 20 local healthy volunteers and 60 patients on long term oral anticoagulant therapy. KC clotting times were plotted against manual clotting times on double logarithmic scales. Orthogonal regression lines were calculated to assess the ISI. RESULTS: In two laboratories the ISI of the KC method was lower than that of the manual method; these differences, however, were 2% or less. In the other laboratories no clinically important differences were observed between ISI values obtained. However, the clotting times determined with the KC methods were shorter than the manual values. CONCLUSIONS: The ISI of Thrombotest determined with the KC methods was very similar to the manual value. Therefore, use of the ISI value supplied by the manufacturer without adjustment is justified. The mean normal prothrombin time, however, must be determined locally.

    Statistics from Altmetric.com

    Request Permissions

    If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.